Temperature control
238
Goldeye G/CL Technical Manual V4.2.0
Goldeye CL-033 TEC1 example
This example explains the functionality of the single-stage TEC1. It is shown in
The Goldeye CL-033 TEC1 has four temperature setpoints:
•
+5 degrees Celsius,
•
+20 degrees Celsius (default),
•
+35 degrees Celsius, and
•
+50 degrees Celsius.
It can achieve a maximum temperature difference of 25 kelvin between housing
and sensor.
As a result, the CL-033 TEC1 switches to the lowest setpoint it can reach, which is
at +35 degrees Celsius, and keeps the sensor at that temperature.
For this example, it is assumed that during operation, the case temperature of a
Goldeye CL-033 TEC1 has leveled at approxi48 degrees Celsius. Thus, the
minimum sensor temperature the TEC can reach is +23 degrees Celsius. It cannot
reach the default setpoint at +20 degrees Celsius.
Goldeye G-032 Cool TEC2 example
This example explains the functionality of the two-stage TEC2.
The Goldeye G-032 Cool TEC2 has three temperature setpoints:
•
-20 degrees Celsius (default),
•
-5 degrees Celsius, and
•
+10 degrees Celsius.
It can achieve a maximum temperature difference of 60 kelvin between housing
and sensor.
For this example, it is assumed that during operation, the case temperature of a
Goldeye G-032 Cool TEC2 has leveled at approxi38 degrees Celsius.
Figure 75: Cooling capabilities of Goldeye TEC1 models (example)
-21 °C
-10 °C
0 °C
10 °C
20 °C
30 °C
40 °C
50 °C
Ambient temperature
Default setpoint (sensor)
Exis
Ɵ
ng setpoint (sensor)
Housing temperature
DŝŶ͘ȴdсϮϱ<
Default setpoint
not reachable
Lowest possible
setpoint: +35 °C