88
STEP 50
CH2 DC offset at
−
full scale.
Check for –10V ± 50mV
STEP 51
CH2 Multiplier zero.
Adjust for minimum Volts AC
STEP 52
CH2 Multiplier offset.
Adjust for 0V ± 3mV
STEP 53
CH2 Waveform offset Filter 1.
Adjust for 0V ± 3mV
STEP 54
CH2 Output level at full
−
scale Filter 1
Adjust for 10V ± 10mV
STEP 55
CH2 Waveform offset Filter 2
Adjust for 0V ± 3mV
STEP 56
CH2 Output level at full
−
scale Filter 2
Adjust for 10V ± 10mV
STEP 57
CH2 20dB attenuator (A)
Adjust for 1V ± 1mV
STEP 58
CH2 20dB attenuator (B)
Adjust for 1V ± 1mV
STEP 59
CH2 10dB attenuator
Adjust for 3.162V ± 3mV
STEP 60
CH2 Sine amplitude flatness
Note reading
STEP 61
CH2 Level 1MHz
Adjust for same reading
STEP 62
CH2 Level 2MHz
Adjust for same reading
STEP 63
CH2 Level 5MHz
Adjust for same reading
STEP 64
CH2 Level 10MHz
Adjust for same reading
STEP 65
CH2 Level 15MHz
Adjust for same reading
STEP 66
CH2 Level 20MHz
Adjust for same reading
STEP 67
CH2 Level 25MHz
Adjust for same reading
STEP 68
CH2 Level 30MHz
Adjust for same reading
STEP 69
CH2 Level 35MHz
Adjust for same reading
STEP 70
CH2 Level 40MHz
Adjust for same reading
STEP 71
CH2 Level 43MHz
Adjust for same reading
STEP 72
CH2 Level 45MHz
Adjust for same reading
STEP 73
CH2 Level 47MHz
Adjust for same reading
STEP 74
CH2 Level 50MHz
Adjust for same reading
STEP 75
CH2 Square-wave high level
Adjust for 10V ± 10mV
STEP 76
CH2 Square-wave low level
Adjust for -10V ± 10mV
STEP 77
CH2 Rise time (0)
Adjust for 50ns
STEP 78
CH2 Rise time (0)
Adjust for 100ns
STEP 79
CH2 Rise time (1)
Adjust for 100ns
STEP 80
CH2 Rise time (1)
Adjust for 2us
STEP 81
CH2 Rise time (2)
Adjust for 2us
STEP 82
CH2 Rise time (2)
Adjust for 40us
STEP 83
CH2 Fall time (0)
Adjust for 50ns
STEP 84
CH2 Fall time (0)
Adjust for 100ns
STEP 85
CH2 Fall time (1)
Adjust for 100ns
STEP 86
CH2 Fall time (1)
Adjust for 2us
STEP 87
CH2 Fall time (2)
Adjust for 2us