the 10MHz REF IN connector will be used, if it is valid, and the Status Line will show
. If the
signal at the 10MHz REF IN connector is invalid or not present the warning message
External
clock invalid Generator running on internal clock
will pop up and the
Status line will show
indicating that the generator is running on internal clock. When a valid
clock is connected the instrument will use it and pop up the message
External clock
detected Generator running on external clock
. If a noisy or otherwise out-of-
specification signal is present at the 10MHz REF IN connector the instrument may alternate
between the two messages.
Pressing the
LokTyp
soft-key gives access to the multi-instrument synchronising capabilities of
the instrument.
See the Synchronising Two Generators section for details of these options.
Calibration
Pressing the
Calib
soft-key shows the opening screen for instrument calibration.
Full details on calibration may be found in the Calibration section.