Running Samples
86
NovoCyte Quanteon Flow Cytometer Operator's Guide
To collect the data, first double-click the sample to be collected in the
Experiment
Manager
to make it the Active Sample. Then place the corresponding sample tube
into the sample holder and click
Run
button. Repeat the procedure until all the
compensation samples are collected.
After all the samples are collected, the NovoExpress software will automatically
generate an FSC/SSC density plot, a histogram and a series of two parameter
density plots with the fluorescence channel of the fluorophore used to stain the
sample as the X-axis parameter and each of the other fluorescence channels as
the Y-axis parameter for each sample. In the FSC-H/SSC-H density plot, a gate
named
Main
will be automatically created for the main compensation beads or cell
populations and be applied to all rest plots. NovoExpress will also set a Bi-Range
gate for the histogram to separate the negative (
Negative
gate) and positive
(
Positive
gate) populations, as shown in Figure 52.
Figure 52.
Data Analysis for PE Single-Stained Sample in Auto Compensation
Under normal circumstances, the automatically generated gates are appropriate.
However, due to the complexity of the sample, in some cases, the automatic
software gating algorithm may not be able to generate the optimal gates. Check
the
Main
,
Negative
, and
Positive
gates to see if they are properly set. When
necessary, adjust the gates manually to optimize the results.
NOTE
If the single-stained sample does not have a negative population, be sure to always
include an
unstained
sample for automatic compensation. Otherwise, the
NovoExpress software cannot conduct automatic compensation and an error
massage will be generated.
NOTE
User may need manually adjust the gates when significant amount of debris or
complex populations are observed for single-stained sample.
4
Automatic Calculation of Compensation Matrix
The NovoExpress software will automatically calculate the compensation matrix
based on the signals (Mean or Median) of the negative and positive populations.
The font color of
Compensation
under the
Compensation Specimen
node in the
Experiment Manager
turns blue to indicate the compensation matrix has non-zero