Instrument Operation
62
NovoCyte Quanteon Flow Cytometer Operator's Guide
current user account by following the instructions described in
Access
Privilege
in
NovoExpress Software Guide
.
2
Click
Instrument
>
Configuration
to open the
Instrument Configuration
window. Click
Create Copy
in the
Instrument Configuration
window to
generate an editable copy of the configuration. Click
Rename
to rename the
configuration if needed. Click
Change Optical Configuration
to continue.
NOTE
A customized optical configuration can be deleted by clicking
Delete
. Only the user-
defined optical configurations can be deleted.
Figure 28.
Customize the Instrument Configuration
3
Open the top cover of the instrument. Insert one end of the Allen wrench
through the hole. Gently hold the proper optical filter or dichroic mirror and pull
it upward to remove it from the slot. Insert the new optical filter or dichroic
mirror back into the slot as shown in the following figure. Close the top cover
of the instrument.