System Maintenance
114
NovoCyte Quanteon Flow Cytometer Operator's Guide
Figure 89.
Software Message Indicating Last Replacement of Fluidic System Consumables is 60 Days
Ago
NOTE
The fluidic system consumable replacement process can also be initiated manually
at any time by clicking the
Instrument
>
Replace Fluidic System Consumables
icon
in the NovoExpress software when the instrument is in
Ready
status.
Figure 90.
Replace Fluidic System Consumables icon
Click the
Replace
button and a dialog box in prompted window below.
Figure 91.
Software Message for Fluidics Consumables Replacement
The following fluidic system consumables should be replaced at the same time:
•
NovoFlow sheath fluid in-line filter (1pcs)
•
Waste filter (1pcs)