System Maintenance
NovoCyte Quanteon Flow Cytometer Operator's Guide
129
NOTE
The
Purge
function is only available when the
Purge Instrument
privilege is
enabled. Refer to
NovoExpress Software Guide
for instructions to enable this
privilege.
1
Click
Instrument > Fluidics Maintenance > Purge
from the NovoExpress
software.
Figure 108.
Purge Function in NovoExpress
2
NovoExpress software will prompt a dialog box to guide you through the
purging procedure. Follow the instructions in each step and click
OK
to move
to the next step until completion.
Figure 109.
Software Message When Starting the Purge Procedure
3
Power off the instrument. Unplug the power supply and accessories.
4
Sterilize the reagent containers following the procedure described in
of this guide.
5
Pack the instrument and accessories properly for shipment.
CAUTION
If there are operating questions when purging the fluidics system, in order to
avoid damaging the instrument, please stop purging and contact your local
Agilent representative or distributor
.