System Maintenance
NovoCyte Quanteon Flow Cytometer Operator's Guide
121
2
Press the metal clip on the quick coupler to disconnect the tubing from the
NovoFlow sheath fluid container' cap.
3
Unscrew the cap and take the tubing-filter assembly (see figure below) out
from the container and put it on a clean lint-free wipe.
Figure 100.
NovoFlow Sheath Fluid Inlet Filter
4
Replace the old filter with a new one and correctly connect the tubing to the
filter.
NOTE
The procedures described above may vary if NovoCyte Fluidics Cart is used.
Please refer to
NovoCyte Fluidics Cart Operator's Guide
for more details.
CAUTION
Please use proper force when tightening the Luer connectors.
CAUTION
The ends of the sheath fluid inlet filter are different to ensure the filter is
connected into the system in the correct direction. Incorrect installation will
lead to system malfunction.
5
Click
Instrument > Fluidics Maintenance > Priming
to prime the fluidic
system.
CAUTION
When replacing the sheath fluid inlet filter, do not twist or kink the tubing.
Replace NovoRinse and NovoClean Solution Inlet Filter
NovoRinse and NovoClean solution inlet filters are placed inside the NovoRinse
and NovoClean containers, respectively, to filter out micro-particles or impurities in
the solution.
1
Make sure the instrument is in
Ready
or shut down status.
2
Press the metal clip on the quick coupler to disconnect the tubing from the
NovoRinse or NovoClean container' cap.