Running Samples
NovoCyte Quanteon Flow Cytometer Operator's Guide
81
Figure 47.
Experiment Control Panel
NOTE
The sample injection probe may perform multiple aspirations for one sampling. Do
not remove or change the test tube until the entire sampling process has
completed.
Note
When
Recover Remaining Samples
is checked, the remaining unused samples will
be automatically returned to the sample tube. Please refer to
Experiment Control
in
NovoExpress Software Guide
for more details on this function.
During the acquisition, the
Cytometer Control
>
Active Sample Information
panel
will show the progress in real-time, as shown in Figure 48.
Figure 48.
Active Sample Information
Once the stop condition is reached, sample acquisition will automatically stop.
NOTE
You can also manually stop the sample acquisition by clicking the
Stop
button in
the
Cytometer Control
>
Experiment Control
panel, during the sample acquisition
process.
After sample acquisition is completed, the NovoCyte Quanteon flow cytometer
automatically executes the cleaning process for the fluidics system and the
sample injection probe. This helps to reduce cross-contamination between
different samples. When the cleaning process is completed, repeat the above
procedure for sample
NBK
.
NOTE
If you want to increase the sampling speed, you can turn off the automatic cleaning
procedure after sampling. To do so, uncheck
Rinse after Sampling
box in the
Cytometer Control
panel from the NovoExpress software as shown in Figure 47.