System Maintenance
124
NovoCyte Quanteon Flow Cytometer Operator's Guide
Figure 103.
Properly Installed 0.05 µm Sheath Fluid Ultrafiltration (UF) Filter
5
Open NovoExpress software. Click
Instrument > Fluidic Maintenance >
Priming
to prime the fluidic system. Ensure no fluidic leakage from the filter’s
inlet/outlet.
6
Load a sample tube with 1 mL Milli-Q or 0.2 µm filtered DI water onto the tube
holder. Create a new experiment. Set
Stop Condition
as 500 μL and
Flow Rate
as 120 μL/min. Click the
Run
button in
Experiment Control
panel to start the
sample acquisition. This step is to flush sample tubing.
7
Close the front panel. The installation is completed, and the instrument is
ready to use.