Maintenance and Troubleshooting
Troubleshooting
Chapter 6
159
Troubleshooting
Use the following table to help troubleshoot your multiport test system.
Troubleshooting the Test System
Symptom
Cause
Cure
One or more biases not
applied.
Bias fuse blown.
Check bias fuses. Replace if
needed with fuse of the
same type and rating.
Control computer can not
communicate with the test
set.
Accidental change to GPIB
switch settings.
Set the GPIB address as
needed. Restart the system.
See
The test set does not come
on the first time you use it.
Line fuse not installed, or
incorrect line fuse installed.
Install the line fuse. Refer to
“How to Set Up the
Multiport Test System
Hardware” on page 18
Excessive ripple in data.
Load termination damaged
by excessive RF power.
Contact Agilent
Technologies. See
“Contacting Agilent” on
page 160
for more
information.
Loose connections between
VNA and test set and/or
between test set and DUT.
Check and torque the
connectors.
Poor test cable repeatability.
See
Cable Repeatability” on
page 75
Replace test cables as
needed. You can replace a
single cable, without
replacing the entire set.
High loss on one path with
poor raw data match (as
seen during analyzer sweep)
or inability to make a good
calibration.
Possible signal channel
damage. See
Test Cable Repeatability” on
page 75
Contact Agilent
Technologies. See
“Contacting Agilent” on
page 160
for more
information.
Summary of Contents for N4413A
Page 4: ...iv ...
Page 11: ...1 1 About the Multiport Test System ...
Page 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Page 19: ...9 2 User Interface ...
Page 27: ...17 3 Using the Multiport Test System ...
Page 89: ...79 4 Performing Time Domain Measurements ...
Page 109: ...99 5 Menu Reference ...
Page 163: ...153 6 Maintenance and Troubleshooting ...
Page 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Page 173: ...163 7 Instrument Information ...
Page 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Page 209: ...199 8 Safety and Regulatory Information ...
Page 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Page 219: ...209 A Other Technical Procedures ...
Page 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...