Performing Time Domain Measurements
Using the Multiport Test System for Time-Domain Measurement and Analysis
Chapter 4
92
Time Domain Controls
Radio buttons allow selection of impulse or step response. When step response is chosen on a
reflection parameter, the data can also be displayed as impedance rather than a reflection
coefficient. Linear (Real) or logarithmic (Log Mag) scales can also be selected.
Fill-in forms allow the time window to be further customized. The units for the start- and
stop-points are specified in either time (nS) or distance (cm), depending on the setting of the
radio button. With distance selected, you may also specify a velocity factor.
Gating Controls
The controls for gating the time domain response provide the ability to remove the effect of a
particular circuit element mathematically. To add a gate, first choose the gate type. An ’Ideal’
gate replaces the gated section with an ideal transmission line having the same electrical
delay as the section that is removed. A ’Zero Length’ gate removes the gated section and
effectively shortens the structure. Then select
Add Gate
.
When
Add Gate
is selected, a dialog box appears to allow the start- and stop-points of the
gate to be defined.
Up to 10 gates can be added. When using multiple gates, ’Ideal’ and ’Zero Length’ gates are
highlighted differently for easy identification. The gates are numbered sequentially from left
to right. For example, to delete the left-most gate, set the
Delete Gate
box to 1 and click on
the
Delete Gate
button. The next gate then becomes gate number 1.
Once a gate has been defined, its position can be adjusted. The start-point can be moved with
the end-point fixed, the end-point can be moved with the start-point fixed, or the segment can
be shifted along the trace. This is accomplished by selecting the gate (by its number), and
clicking on the right or left soft arrow keys (not the keyboard).
By observing the original frequency domain response and the transformed frequency domain
response, the effect of the gating operation on the S-parameter data can be seen.
Summary of Contents for N4413A
Page 4: ...iv ...
Page 11: ...1 1 About the Multiport Test System ...
Page 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Page 19: ...9 2 User Interface ...
Page 27: ...17 3 Using the Multiport Test System ...
Page 89: ...79 4 Performing Time Domain Measurements ...
Page 109: ...99 5 Menu Reference ...
Page 163: ...153 6 Maintenance and Troubleshooting ...
Page 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Page 173: ...163 7 Instrument Information ...
Page 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Page 209: ...199 8 Safety and Regulatory Information ...
Page 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Page 219: ...209 A Other Technical Procedures ...
Page 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...