Using the Multiport Test System
How to Characterize Adapters
Chapter 3
45
3. Select the appropriate adapter characterization category (
“In-Series, Same-Sex”
or
“All
Others”
) and calibration kit(s).
4. Follow the program prompts to perform the characterization.
When the characterization is complete, save the adapter file using the default
.TXS
file
extension. Since the characterization file is in Citifile format, you can import the data and
make a visual check of the quality of the characterization. Select
File > Import > Citifile >
and open and inspect the adapter characterization file.
One-Port Adapter Characterization
Two-Port Adapter Characterization
See
“Setup > De-Embedding…” on page 115
and
“Setup > Modify Rotations…” on page 119
more information on characterization and de-embedding.
Summary of Contents for N4413A
Page 4: ...iv ...
Page 11: ...1 1 About the Multiport Test System ...
Page 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Page 19: ...9 2 User Interface ...
Page 27: ...17 3 Using the Multiport Test System ...
Page 89: ...79 4 Performing Time Domain Measurements ...
Page 109: ...99 5 Menu Reference ...
Page 163: ...153 6 Maintenance and Troubleshooting ...
Page 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Page 173: ...163 7 Instrument Information ...
Page 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Page 209: ...199 8 Safety and Regulatory Information ...
Page 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Page 219: ...209 A Other Technical Procedures ...
Page 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...