Maintenance and Troubleshooting
Care of Test Cable Assemblies
Chapter 6
156
Depending on the connector, over-torque can cause damage to connectors in a variety of ways:
• mushroomed outer interface shells
• mushroomed pin shoulders
• recessed or protruding pins
• recessed or protruding dielectrics
• bent pins
• chipped plating
• coupling nut retaining ring damage
• damage to coupling threads
Cable Handling
The test cables have a 1 inch minimum bend radius. Exceeding this radius can result in poor
measurements and poor repeatability. Be alert to tight bends where they are not necessarily
obvious — like at the end of connector strain relief tubing.
Swept 90
°
adapters may be used (typically, at the test set front panel) to minimize cable
bending. However, if the adapters are used, they must be in place during the calibration.
Cables are often stored in a coiled configuration. Coiled cable “set” (the tendency to stay
coiled) can occur if the cables are left coiled. Use large coil diameters (one or two feet) to
minimize cable set. Unroll coiled cables prior to use — never just pull out the loops.
Avoid pinching, crushing, or dropping objects on cable assemblies. Dragging a cable over a
sharp edge will tend to flatten one side, and it is highly likely that the minimum bend radius
will be exceeded.
Summary of Contents for N4413A
Page 4: ...iv ...
Page 11: ...1 1 About the Multiport Test System ...
Page 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Page 19: ...9 2 User Interface ...
Page 27: ...17 3 Using the Multiport Test System ...
Page 89: ...79 4 Performing Time Domain Measurements ...
Page 109: ...99 5 Menu Reference ...
Page 163: ...153 6 Maintenance and Troubleshooting ...
Page 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Page 173: ...163 7 Instrument Information ...
Page 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Page 209: ...199 8 Safety and Regulatory Information ...
Page 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Page 219: ...209 A Other Technical Procedures ...
Page 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...