Using the Multiport Test System
How to Perform Electronic Calibration
Chapter 3
43
3. Modify any parameters in the
Measurement Parameters
dialog box as required and click
OK
.
4. When prompted, select the module characterization file for your ECal module. This file is
located on the floppy disk provided with the module. Each module characterization file is
unique, so verify that the module serial number matches the file’s serial number.
5. A progress meter is displayed during the calibration. When complete, assign the
calibration a file name and save the calibration data.
NOTE
Creating a Calibration Subset
If you do not want to use all frequency points for your measurement, you can
create a calibration subset and use it for the active calibration instead of the
full calibration set. See
“Calibrate > Create Cal Subset…” on page 133
.
Summary of Contents for N4413A
Page 4: ...iv ...
Page 11: ...1 1 About the Multiport Test System ...
Page 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Page 19: ...9 2 User Interface ...
Page 27: ...17 3 Using the Multiport Test System ...
Page 89: ...79 4 Performing Time Domain Measurements ...
Page 109: ...99 5 Menu Reference ...
Page 163: ...153 6 Maintenance and Troubleshooting ...
Page 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Page 173: ...163 7 Instrument Information ...
Page 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Page 209: ...199 8 Safety and Regulatory Information ...
Page 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Page 219: ...209 A Other Technical Procedures ...
Page 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...