Using the Multiport Test System
How to Use Frequency Zoom
Chapter 3
56
Increasing Frequency Step
The multiport test system allows measurements to be made at frequency step sizes larger
than the step size used in the calibration. This is accomplished by skipping points.
To skip points, follow these steps:
1. Right click in the gray area outside the plot to open the display properties menu.
2. In the display properties menu, select
Define Point Skip
. The
Select Number of Points to
Skip
dialog box appears:
3. Select the number of points (0 to 10) to skip between two measured points.
4. Select one of the buttons at the bottom of the dialog to finish the selection.
5. To restore the original measurement condition, right click in the gray area outside of the
plot, and select
Restore Full Sweep Data
from the menu.
If the frequency step has been increased, the
+Data
button appears in the indicator bar.
If a measurement is made after increasing the frequency step, fewer points are measured.
This creates a new set of calibration data adjusted to the new frequency step.
As soon as the measurement starts, the
+Cal
button appears in the indicator bar to show that
the calibration in use is a subset of the full calibration available.
At the end of the measurement, the
+Data
button disappears because all available
measurement point have been displaced.
APPLY
Applies the new settings without closing the dialog box.
OK
Applies the new settings and closes the dialog box.
Cancel
Closes the dialog box without making any changes.
Summary of Contents for N4413A
Page 4: ...iv ...
Page 11: ...1 1 About the Multiport Test System ...
Page 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Page 19: ...9 2 User Interface ...
Page 27: ...17 3 Using the Multiport Test System ...
Page 89: ...79 4 Performing Time Domain Measurements ...
Page 109: ...99 5 Menu Reference ...
Page 163: ...153 6 Maintenance and Troubleshooting ...
Page 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Page 173: ...163 7 Instrument Information ...
Page 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Page 209: ...199 8 Safety and Regulatory Information ...
Page 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Page 219: ...209 A Other Technical Procedures ...
Page 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...