Menu Reference
Setup Menu
Chapter 5
113
Select the Calibration Instruction Set
allows tailoring of the sequence for presenting
calibration standards to the system for the best match with the measurement environment.
Setup > Adapter Usage…
Pull down
Setup > Adapter Usage…
to open the
THRU ADAPTER USAGE
dialog box and
setup adapters for use in calibrating the system.
Click the
Select Thru Configuration
choice that best minimizes the amount of cable
bending during the calibration and the measurement. The choices are:
•
Config 1
- uses four paths (1-2, 3-4, 1-4, 3-2)
•
Config 2
- uses four paths (1-2, 3-4, 1-3, 2-4)
•
Config 3
- uses all six paths (1-2, 3-4, 1-3, 2-4, 1-4, 3-2)