Performing Time Domain Measurements
General Theory
Chapter 4
81
An alternative method of obtaining time-domain characterization of a device is to make the
measurement directly in the time domain by synthesizing a step waveform, applying it to the
device, and observing the response on an oscilloscope. The advantages of using the
measurement approach for TDR data are listed below.
While the traditional TDR measurement technique provides fast measurement speed, the
measurement technique used by the multiport test system provides:
• Superior accuracy
• Significantly better dynamic range (important for crosstalk and mode-conversion terms)
• Ability to de-embed fixtures and signal launchers
• Access to both frequency and time domain information (as vector quantities)
• Single setup for forward and reverse transmission and reflection, single-ended,
differential-, and common-mode, and mode-conversion terms
• No need for DUT to have DC return path
• No large voltage steps applied to DUT
Summary of Contents for N4413A
Page 4: ...iv ...
Page 11: ...1 1 About the Multiport Test System ...
Page 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Page 19: ...9 2 User Interface ...
Page 27: ...17 3 Using the Multiport Test System ...
Page 89: ...79 4 Performing Time Domain Measurements ...
Page 109: ...99 5 Menu Reference ...
Page 163: ...153 6 Maintenance and Troubleshooting ...
Page 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Page 173: ...163 7 Instrument Information ...
Page 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Page 209: ...199 8 Safety and Regulatory Information ...
Page 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Page 219: ...209 A Other Technical Procedures ...
Page 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...