Using the Multiport Test System
How to Set Up the Multiport Test System Hardware
Chapter 3
38
To Set the GPIB Addresses
The test set uses GPIB to communicate with system components. The GPIB address switch is
on the back panel of the test set; it is set to “18” at the factory. While this will work in most
situations, the address can be changed if required to avoid address conflicts with other
equipment on the same bus. If you change the address on the test set, make sure you also
change it in the software. (See
“Setup > Hardware…” on page 124
GPIB addresses are five-bit binary numbers set with the switches labeled “16” through “1”,
with “1” being the least significant digit.
“T”, “L”, and “S” correspond to Talk, Listen, and Status. Set them as shown (off).
The example below shows the default setting, “18”, set by turning on switches “16” and “2”.
NOTE
After changing a GPIB address setting, on any of the system equipment, turn
the test set AC power off and back on establish the new GPIB address.
Pull down
Setup > Hardware…
to set the test set GPIB address in the software. See
Summary of Contents for N4413A
Page 4: ...iv ...
Page 11: ...1 1 About the Multiport Test System ...
Page 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Page 19: ...9 2 User Interface ...
Page 27: ...17 3 Using the Multiport Test System ...
Page 89: ...79 4 Performing Time Domain Measurements ...
Page 109: ...99 5 Menu Reference ...
Page 163: ...153 6 Maintenance and Troubleshooting ...
Page 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Page 173: ...163 7 Instrument Information ...
Page 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Page 209: ...199 8 Safety and Regulatory Information ...
Page 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Page 219: ...209 A Other Technical Procedures ...
Page 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...