Agilent InfinityLab LC Series Vialsamplers User Manual
99
3
Using the Module
Using the Optional Integrated Column Compartment
Method Parameters
The
Method
dialog box enables the user to change the method parameters
relevant for the ICC:
• Temperature
:
•
Not Controlled
: The temperature of the column oven stays uncontrolled.
•
At specified temperature (
°C
): The temperature of the column oven is
regulated to stay at the setpoint temperature specified.
•
Unchanged
: The temperature setting stays unchanged, meaning that the
setting currently stored in the system is used by the method.
• Valve Position/Column
: This section provides information about the location
of the column that is being used for the given method.
•
Enforce column for run
: Marking this checkbox restricts the execution of
the method to a specific column type, identified by its product number.
• Stoptime
:
•
As Pump/Injector
: The ICC uses the same stoptime setting as specified for
the pump or the sampler.
•
After specified time (
min
): The stop time defines a time limit for the
analysis, after which all method parameters are reset to initial values.
• Posttime
:
•
Off
: No post-run period is applied.
•
For specified time (
min
): After the analysis, the system stays in the
post-run state for the specified time. Using a post-time period can be
beneficial when the column needs to be re-equilibrated before starting the
next analysis (for example, gradient elutions).
NOTE
Valid inputs are 10 to 80 °C in increments of 0.1 °C.
NOTE
For
Stoptime
and
Posttime
, valid inputs are 0.01 to 99999 minutes in increments
of 0.1 min.