Agilent InfinityLab LC Series Vialsamplers User Manual
76
3
Using the Module
Install the Optional Integrated Column Compartment
Install the Integrated Column Compartment
For more information on recommended capillary types refer to
Tools required
p/n
Description
8710-0510
Open-end wrench 1/4 — 5/16 inch
Parts required
#
p/n
Description
1
G7130-60030
Integrated Column Compartment, 3 µL Heater
OR
1
G7130-60060
Integrated Column Compartment, 6 µL Heater
1
Accessory Kit for ICC
(includes capillaries and column holder clips)
Preparations
•
Switch off the pump.
•
Switch off the sampler.
•
Remove the column.
•
Open the doors of the sampler.
WARNING
Fire or explosion due to missing leak drainage
Solvents may run into the module if column shelf or Integrated Column
Compartment are not installed.
Install either column shelf or Integrated Column Compartment before
starting the module.