Agilent InfinityLab LC Series Vialsamplers User Manual
240
8
Maintenance
Exchange the Needle Seat Assembly
17
In Lab Advisor, select
Next
to move the needle into the
needle seat and then
Back
to finish the maintenance
procedure.
OR
In the Local Controller, select
Continue
to move the nee-
dle into the needle seat and then
Exit
to finish the main-
tenance procedure.
18
Close the doors.
19
Perform a
System Pressure Test
.