Agilent InfinityLab LC Series Vialsamplers User Manual
264
8
Maintenance
Exchange the Wash Port Assembly
3
Verify that the needle in the uppermost position before
continuing.
4
Loosen the safety cover screw.
WARNING
Sharp needle
Uncovered needles may cause injuries
Do not touch the tip of the needle.
5
Remove the safety cover.
WARNING
Sharp needle
Uncovered needles may cause injuries.
Do not change the needle seat at this point.
Do not touch the tip of the needle.
6
Move the wash port into the service position by slightly
lifting it up and then flapping to the left.