Using the Module
Using the Optional Sample Cooler/Sample Thermostat
Frequent Door/Drawer Opening
Opening the door(s) and/or the sample drawers frequently can compromise the
temperature stability, as fresh warm and humid air will enter each time. In a
highly humid environment, this could also lead to the formation of significant
amounts of condensate on the internal surfaces of the sampler.
Ice Formation
The Sample Cooler/Sample Thermostat was designed to operate without the risk
of icing. In an unlikely event of ice formation, turn off the cooler/thermostat and
wait until it defrosts.
Shutting Down
When the Sample Cooler/Sample Thermostat needs to be turned off for the night
or a longer period, the following best practices are recommended:
•
Remove all sample containers and/or vials from the sampler.
•
Let the system reach the ambient temperature. Opening the door(s) of the
sampler facilitates this process.
•
Remove any condensate that might appear on the sample drawers or the
internal surfaces of the sampler.
•
Make sure that all condensate is removed from the cooler/thermostat.
NOTE
Do not use mechanical devices or other means to accelerate the defrosting
process.
NOTE
Gently tapping on the sides of the sampler facilitates the condensate removal.
Tilting the module towards its right back corner is not recommended as it can
damage the internal parts.