Exchange the Gripper Arm
Exchange the Gripper Arm
When
The gripper arm needs to be replaced due to malfunctioning.
Tools required
p/n
Description
5023-2499
Hex Key Set
(The 1.5 mm hex key is needed)
Parts required
p/n
Description
G1313-60010
Gripper assembly
Preparations
Finish any pending acquisition job
NOTE
Alternatively, any other tool or item can be used that can be fitted into the block
position slot of the transport arm (for example, a straightened paper clip, or a
small-sized screwdriver).
1
In the Agilent Lab Advisor software, select
Service &
Diagnostics> Maintenance Positions> Change Gripper
and wait until the gripper is in maintenance position
(see also
OR
In the Local Controller start the maintenance mode and
select
Change Gripper
function.
2
Open the doors of the module.