Error Information
Vialsampler Error Messages
Vialsampler Error Messages
Command failed errors
These are non-specific error messages and only indicate that something went
wrong during the execution of the given action command. They are usually
coupled with other error events, which provide more specific information on the
nature of the issue. To identify these source error events, look for entries with the
same timestamp as for the
Command failed
error message.
•
Error ID: 34202
→
Metering command failed
•
Error ID: 34205
→
Draw command failed
•
Error ID: 34208
→
Eject command failed
•
Error ID: 34211
→
Mix command failed
•
Error ID: 34214
→
Needle command failed
•
Error ID: 34217
→
Transport command failed
•
Error ID: 34220
→
Valve command failed
Probable cause
Suggested actions
1
Command failed
Identify the source error message and follow the
instructions provided for that.