![Agilent Technologies G7129A User Manual Download Page 50](http://html1.mh-extra.com/html/agilent-technologies/g7129a/g7129a_user-manual_2867753050.webp)
Agilent InfinityLab LC Series Vialsamplers User Manual
50
2
Site Requirements and Specifications
Specifications of the Sample Thermostat
CAUTION
General hazards and improper disposal
Improper disposal of the media and components used pollutes the
environment.
The disposal or scrapping of the Sample Cooler or the Sample Thermostat
must be carried out by a qualified disposal company.
All media must be disposed of in accordance with national and local
regulations.
Please contact your local Agilent Service Center in regard to safe
environmental disposal of the appliance or check
www.agilent.com
for
more info.
Table 8
Performance Specifications for the Sample Thermostat
Type
Specifications
Operating principle
High performance, low-energy consumption micro-compressor based
cooler with natural R600a coolant (Butane 30 g), user-upgradable
Temperature range
from 4 – 40 °C
Temperature settable
from 4 – 40 °C in 1 ° increments
Temperature accuracy
(
<
25 °C,
<
50 % r.H.)
2 °C to 6 °C at a setpoint of 4 °C
NOTE
Minimum firmware revision for the Sample Thermostat is D.07.22.
Minimum LC driver revision for the Sample Thermostat is A.02.14.