Agilent InfinityLab LC Series Vialsamplers User Manual
219
7
Error Information
Integrated Column Compartment (ICC) Heater Error Messages
Integrated Column Compartment (ICC) Heater
Error Messages
Wait for temperature function timed out
Error ID: 30820
The system reported an error because reaching the set temperature took longer
than expected. To verify the correct functioning of the heating, perform the
Heater Test in the Agilent Lab Advisor software.
Heater over-current protection active
Error ID: 30821
The heating function of the Integrated Column Compartment is switched due to
overcurrent.
To verify the correct functioning of the heating element, perform the
Heater Test
in the Agilent Lab Advisor software.
Probable cause
Suggested actions
1
Connection problems
•
Power off the system and verify that the ICC
is correctly mounted in the sampler.
•
Power on the system and check if the prob-
lem persists.
2
Potential hardware error
Please contact your Agilent service representa-
tive.
Probable cause
Suggested actions
1
Connection problems
•
Power off the system and verify that the ICC
is correctly mounted in the sampler.
•
Power on the system and check if the prob-
lem persists.
2
Potential hardware error
Please contact your Agilent service representa-
tive.