Error Information
Integrated Column Compartment (ICC) Heater Error Messages
A heater sensor reading failed
Error ID: 30822
The heating function of the Integrated Column Compartment is switched off due
to the malfunctioning of the digital temperature sensor.
To verify the correct functioning of the heating element, perform the
Heater Test
in the Agilent Lab Advisor software.
Trend analysis for the heater temperature sensors
failed
Error ID: 30823
The heating function of the Integrated Column Compartment is switched off
because the trend analysis of the reading from the digital temperature sensor
has failed.
To verify the correct functioning of the heating element, perform the
Heater Test
in the Agilent Lab Advisor software.
Probable cause
Suggested actions
1
Connection problems
•
Power off the system and verify that the ICC
is correctly mounted in the sampler.
•
Power on the system and check if the prob-
lem persists.
2
Potential hardware error
Please contact your Agilent service representa-
tive.
Probable cause
Suggested actions
1
Connection problems
•
Power off the system and verify that the ICC
is correctly mounted in the sampler.
•
Power on the system and check if the prob-
lem persists.
2
Potential hardware error
Please contact your Agilent service representa-
tive.