Agilent InfinityLab LC Series Vialsamplers User Manual
273
8
Maintenance
Replace the Sample Cooler/Sample Thermostat
7
Slide the new cooler/thermostat halfway into the sam-
pler and connect the power and the data cable.
CAUTION
Damage to the cables
Do not bend or pinch the cables.
Make sure that the Sample Cooler/Sample Thermostat
fits perfectly in the sampler.
8
Slide the cooler/thermostat all the way into the sampler,
making sure that the cables don’t get jammed between
the metal parts.
9
Fix the unit with the four screws.
10
Reconnect the condensate tubing.
NOTE
For information on proper condensate handling, see