Agilent InfinityLab LC Series Vialsamplers User Manual
216
7
Error Information
Sample Cooler/Sample Thermostat Error Messages
Cooler condenser fan failed
Error ID: 30726
The condenser fan of the cooler/thermostat is not working properly.
Thermostat communication error
Error ID: 30738
The system is in the error state because the communication between the
sampler and the thermostat has failed.
Heater defect
Error ID: 30739
The heating function of the thermostat is not working properly.
Probable cause
Suggested actions
1
Potential hardware error
Please contact your Agilent service representa-
tive.
Probable cause
Suggested actions
1
Potential hardware error
Please contact your Agilent service representa-
tive.
Probable cause
Suggested actions
1
Potential hardware error
Please contact your Agilent service representa-
tive.