Agilent G3437A Side-Mounted Thermal Conductivity Detector
Installation Instructions
11
Attach the third detector flow frame
1
Connect the third detector heated zone cable (attached to the back of the
third detector flow frame) to the MSD heated zone cable (long), which was
attached to the GC in a previous step.
2
Orient the third detector flow frame with the GC by positioning the circular
cutouts at the bottom edge of the frame with the four screw holes at bottom
edge of the GC.
3
Using a Torx driver, attach the frame first with two 4 mm × 12 captive
screws at the top edge of the frame and then with four screws through the
circular cut outs at the bottom edge of the frame.
Figure 14
Third detector flow frame mounted to GC
Top screw locations
Bottom screw
locations