14
Installation Instructions
Agilent G3437A Side-Mounted Thermal Conductivity Detector
7
Orient the pneumatics tubing downwards so that it doesn’t interfere with
the TCD board installation.
Figure 18
TCD module installed
Attach the T-manifold assembly
1
Using a 1/8” nut and ferrule, attach one end of the T-manifold assembly
short tubing arms to the EPC module reference gas fitting.
2
Using a 1/8” nut and ferrule, attach one end of the T-manifold assembly
short tubing arms to the EPC module makeup gas fitting.
Pneumatic tubing
Heater cable connected and
tucked behind bezel
Screw locations