Agilent G3437A Side-Mounted Thermal Conductivity Detector
Installation Instructions
19
7
Connect the Third DET cable to the TCD board in the bottom connector.
Figure 25
TCD module cable connections
Restore the GC to operating condition
1
If installed, remove the red TCD cap on the TCD exhaust.
2
Install the third detector side cover using two Torx flathead screws.
3
Replace the top and side panels.
4
Plug in the GC and turn on the power.
5
Unlock the GC keypad.
a
Press
[Options]
.
b
Select
Keyboard & Display
.
c
Select
Hard Configuration lock
, then press
[No/Off]
.
6
On the GC keypad, press
[Configure]
then
[Aux Det #]
.
7
Select the Aux Detector and press
[Enter]
.
8
On the keypad, press
[Mode/Type]
.
9
Scroll to select the Aux Heater and press
[Enter]
. The GC displays a prompt
to select a heater to install. Select
Aux 1
if you attached the MSD heated
zone cable to plug A1 and select
Aux 2
if you used plug A2. A caution
message will appear instructing you to reboot.
Switching
valve
connections
Third DET
connection