Agilent G3437A Side-Mounted Thermal Conductivity Detector
Installation Instructions
9
4
Attach the two blank label plates to the front side of the bezel to completely
cover the holes.
Figure 10
Attaching the two blank label plates
5
Attach the third detector heated zone cable (short) to the frame by
inserting the cable plug into the sheet metal notch, inserting the cable into
the grommet, and sliding the grommet into the cutout on the frame.
Figure 11
Third detector heated zone cable installed in frame
6
Attach the bulkhead fitting retainer to the rear facing side of the frame by
tightening the screw. See
Figure 12
.
7
Insert the bulkhead fitting through the frame so that the fixed nut is
positioned in the retainer.
Blank label plate
Front side of bezel
Grommet
Third detector heated zone
cable
Sheet metal notch