Agilent Technologies
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warran-
ties, either express or implied, with regard to this manual and any information contained herein, including
but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall
not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or
performance of this document or of any information contained herein. Should Agilent and the user have a
separate written agreement with warranty terms covering the material in this document that conflict with
these terms, the warranty terms in the separate agreement shall control.
©
Agilent Technologies, Inc. 2007
Printed in USA and China
Second edition, June 2007
Agilent Technologies, Inc.
2850 Centerville Road
Wilmington, DE 19808-1610 USA
10
Reboot the GC.
a
Press
[Options]
.
b
Scroll to
Communications
and press
[Enter]
.
c
Scroll to
Reboot the GC?
and press
[On/Yes]
twice to reboot the GC and
have the changes take effect.
11
The new EPC you have installed
must
be properly calibrated before the
detector is ready for application use. To calibrate the pressure and gas type,
refer to the Calibration section of the Advanced User Guide.
12
After completing pressure and gas type configuration, connect the carrier
gas source line to the bulkhead fitting, restore gas pressure, and leak check
all fittings.
13
To calibrate the flow, refer to the Calibration section of the Advanced User
Guide.