Agilent G3437A Side-Mounted Thermal Conductivity Detector
Installation Instructions
13
Install the TCD module
1
Temporarily rest the detector module on the floor of the third detector flow
frame. Slide the bottom cage of the EPC module into the supporting slots on
the frame.
2
Tilt the EPC module forward slightly and position the switching valve wire
in the frame cut out slot, making sure that it is not pinched when the
module is in place against the instrument. See
Figure 17
.
3
Attach the EPC module to the frame at the two upper screws.
Figure 17
Installed EPC module
4
Place the detector over the circular detector cut out.
5
Attach the detector to the GC by using three Torx screws.
6
Connect the TCD heater cable to the third detector heated zone cable plug
located on the third detector flow frame wall above the detector. Tuck the
cable between the side of the GC and the bezel. See
Figure 18
.
Switching valve wire
not pinched
Screws
Supporting slots
EPC module