72
Agilent E8462A Relay Multiplexer Command Reference
ARM
The
ARM
subsystem selects the number of scanning cycles (1 to 32767) for each
INITiate
command.
Subsystem Syntax
ARM
:COUNt <
number
> MIN|MAX
:COUNt? [MIN|MAX]
ARM:COUNT
ARM:COUNt <
number
> MIN|MAX
Allows scanning cycles to occur a multiple of
times (1 to 32,767) with one
INITiate
command when
INITiate:CONTinuous OFF|0
is set.
MIN
sets 1 cycle and
MAX
sets 32,767 cycles.
Parameters
Comments
Number of Scans:
Use only numeric values between 1 and 32767,
MIN
, or
MAX
for
the number of scanning cycles.
Related Commands:
ABORt
,
INITiate:IMMediate
*RST Condition:
ARM:COUNt 1
Example
Setting Ten Scanning Cycles
This example sets a multiplexer module for 10 scans of channels 10 through 17 in a
single-module switchbox.
ARM:COUN 10
10 scans per INIT command.
SCAN (@10010:10017)
Scan channels 10 to 17.
INIT
Start scan, close channel 10.
ARM:COUNt?
ARM:COUNt? [MIN|MAX]
Returns the current number of scanning cycles set by
ARM:COUNt
. The current number of scan cycles is returned when
MIN
or
MAX
is
not specified. With
MIN
or
MAX
as a parameter,
MIN
returns 1 and
MAX
returns
32,767.
Parameter
Name
Parameter
Type
Range of Values
Default
Value
number
numeric
1 thru 32,767|MIN|MAX
1
Summary of Contents for E8462A
Page 2: ...x...
Page 10: ...8 Notes...
Page 11: ...9 Notes...
Page 12: ...10 Notes...
Page 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Page 68: ...66 Using the Multiplexer...
Page 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Page 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Page 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Page 146: ...144 Register Based Programming...
Page 156: ...154 Index...