Agilent E8462A Scanning Voltmeter Application Examples
119
Chapter 4
.
The Scanning Voltmeter
The switchbox device driver discussed in Chapter 2 DOES NOT apply to the
E8462A multiplexer when part of a scanning voltmeter. The note on Page
117 prescribes the use of the Agilent E1326/E1411 5½-Digit Multimeter
device driver and command reference. The multimeter's Command Quick
Reference is provided at the end of this chapter for your scanning voltmeter
application reference. Use your Agilent E1326/E1411 5½-Digit Multimeter
User's Manual for detailed information about multimeter commands used
with the scanning voltmeter. The scanning voltmeter is easily configured to
perform voltage, 2-wire ohms, 4-wire ohms or a combination of 1-, 2-, 3- or
4-wire measurements over the module's internal analog bus. The E1326/
E1411 multimeter automatically controls the channels and tree relays when
you use the
MEASure
or
CONFigure
commands. Tree relays are described
in Table 4-2.
Table 4-1.
Agilent E8462A Default Conditions for Power-on and Reset
Parameter
Default
Value
Description
ARM:COUNt
1
Number of scanning cycles is one.
TRIGger:SOURce
IMM
Advances through a scanning list
automatically.
INITiate:CONTinuous
OFF
Number of scanning cycles is set by
ARM:COUNt.
OUTPut[:STATe]
OFF
Trigger output from EXT, TTL, or ECL
sources is disabled.
[ROUTe:]SCAN:MODE
NONE
Channel list is set up for volts
measurement (the default mode).
[ROUTe:]SCAN:PORT
NONE
Analog bus connections are
disabled.
Channel state
All 256 channels are open (channels 00 - 255)
Tree relay state
All tree relays are open (channels 9000 - 9108)
Analog Bus relay state
All analog bus relays are open
(channels 9200 - 9204)
Channel list from SCAN
command (after *RST)
Current channel list is invalidated following a
reset of the module with *RST command.
Summary of Contents for E8462A
Page 2: ...x...
Page 10: ...8 Notes...
Page 11: ...9 Notes...
Page 12: ...10 Notes...
Page 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Page 68: ...66 Using the Multiplexer...
Page 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Page 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Page 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Page 146: ...144 Register Based Programming...
Page 156: ...154 Index...