Agilent E8462A Relay Multiplexer Command Reference
INITiate:CONTinuous?
INITiate:CONTinuous?
Queries the scanning state. With continuous scanning
enabled, the command returns "1" (ON). With continuous scanning disabled, the
command returns "0" (OFF).
Example
Query Continuous Scanning State
This example enables continuous scanning of a switchbox and queries the state.
Since continuous scanning is enabled,
INIT:CONT?
returns "1".
INIT:CONT ON
Enable continuous scanning.
INIT:CONT?
Query continuous scanning state.
INITiate[:IMMediate]
INITiate[:IMMediate]
Starts the scanning process and closes the first channel in the
channel list. Successive triggers from the source specified by the
TRIGger:SOURce
command advance the scan through the channel list.
Comments
Starting the Scanning Cycle:
The
INITiate:IMMediate
command starts scanning by
closing the first channel in the channel list. Each trigger received advances the scan
to the next channel in the channel list. An invalid channel list definition causes an
error (see
[ROUTe:]SCAN
on page 96).
Stopping Scanning Cycles:
See the
ABORt
command on page page 71.
Example
Enabling a Single Scan
This example enables a single scan of channels 50 through 57 of a single-module
switchbox. The trigger source to advance the scan is immediate (internal) triggering
set with
TRIGger:SOURce IMMediate
(default).
SCAN (@150:157)
Scan channels 50 to 57.
INIT
Begin scan, close channel 50 (use
immediate triggering).
Summary of Contents for E8462A
Page 2: ...x...
Page 10: ...8 Notes...
Page 11: ...9 Notes...
Page 12: ...10 Notes...
Page 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Page 68: ...66 Using the Multiplexer...
Page 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Page 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Page 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Page 146: ...144 Register Based Programming...
Page 156: ...154 Index...