Agilent E8462A Relay Multiplexer Command Reference
109
Example
Cycle through a self-test and return the number of times it failed.
TEST:NUMB?
110,5
Cycle through self-test 10 five times on
card number 1 (test_number parameter is
110) and return the number of times the
self-test fails.
Test all relays on the module.
SCAN:PORT
ABUS
Set the module port to analog bus.
FUNC
2,WIRE1X8
Set module number 2 to the 1-wire by 8
MUX mode.
TEST:NUMB?
299,4
Cycle through self-test 99 four times on
card number 2 (test_number parameter is
299) and return the number of times the
self-test fails. This test operates all relays
on the card.
cc99
This is a special test that scans through all relays.
DO NOT
USE this test when the card has terminal block connections.
NOTE:
The Multiplexer function must be set to the eight 1-wire
MUX mode (
[ROUTe:]FUNC
cc,WIRE1X8
where cc is the
card number) to test relays C100 through C108. Otherwise,
these relays are not tested. This is a precaution taken that
prevents shorting of field wiring if terminal block connections
are inadvertently left connected.
This test first opens all the relays, then closes one relay at a time.
Register read-back is used to determine if all bits of all relay
registers are independent. Then it restores the original relay
pattern. This test is not performed during a *TST? self-test since
the test disturbs the switch connections.
test_number
Description
Summary of Contents for E8462A
Page 2: ...x...
Page 10: ...8 Notes...
Page 11: ...9 Notes...
Page 12: ...10 Notes...
Page 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Page 68: ...66 Using the Multiplexer...
Page 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Page 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Page 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Page 146: ...144 Register Based Programming...
Page 156: ...154 Index...