114
Agilent E8462A Relay Multiplexer Command Reference
IEEE 488.2 Common Command Reference
The following table lists the IEEE 488.2 Common (*) Commands that apply to the
Agilent E8462A module. The operation of some of these commands is described
earlier in this manual. For more information on Common Commands, refer to the
Agilent E1406A Command Module User's Manual
or the ANSI/IEEE Standard
488.2-1987.
Table 3-2.
Command
Command Description
*CLS
Clears all status registers (see STATus:OPERation[:EVENt]?) and clears the error queue.
*ESE
<register
value>
Enable Standard Event.
*ESE?
Enable Standard Event Query.
*IDN?
Instrument ID Query; returns identification string of the module.
*OPC
Operation Complete.
*OPC?
Operation Complete Query.
*RCL
<numeric
state>
Recalls the instrument state saved by *SAV. You must reconfigure the scan list.
*RST
Resets the module. Opens all channels and invalidates current channel list for scanning. Sets
ARM:COUN 1, TRIG:SOUR IMM, and INIT:CONT OFF, DIAG:SCAN:DEL 0.0.
*SAV
<numeric
state>
Stores the instrument state but does not save the scan list.
*SRE
<register
value>
Service request enable, enables status register bits.
*SRE?
Service request enable query.
*STB?
Read status byte query.
*TRG
Triggers the module to advance the scan when scan is enabled and trigger source is
TRIGger:SOURce BUS.
*TST?
Self-test. Executes an internal self-test and returns only the first error encountered. Does not
return multiple errors. The following is a list of responses you can obtain where “cc” is the card
number with the leading zero deleted.
+0 if self test passes.
+cc01 for firmware error.
+cc02 for bus error (problem communicating with the module).
+cc03 for incorrect ID information read back from the module's ID register.
+cc10 if an interrupt was expected but not received.
+cc11 if the busy bit was not held for a sufficient amount of time.
+cc14 if one of the four bank fuses is open.
+cc15 if the data read from a register is different from the register’s setting.
*WAI
Wait to Complete.
Summary of Contents for E8462A
Page 2: ...x...
Page 10: ...8 Notes...
Page 11: ...9 Notes...
Page 12: ...10 Notes...
Page 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Page 68: ...66 Using the Multiplexer...
Page 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Page 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Page 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Page 146: ...144 Register Based Programming...
Page 156: ...154 Index...