32
Configuring the Agilent E8462A Multiplexer
Initial Operation
You must download the Agilent E8462A SCPI driver into the Agilent
E1405/E1406 Command Module to perform the initial operation.
At power-on or following a reset of the module (
*RST
command), all 256
channels are open. A
*RST
command invalidates the current scan list (that
is, you must specify a new scan list). Command parameters are set to the
default conditions as shown below.
Execute
SCAN:PORT ABUS
to enable use of the analog bus for the
SCAN
command. The SCPI driver will then automatically open and close the five
analog bus relays during a scan. A
CLOSe
command on a channel will also
automatically close the appropriate tree relays for the given
ROUTe:FUNCtion (see Page 13 "Figure 1-1. Agilent E8462A Simplified
Schematic" ).
Note
Do not execute register writes if you are controlling the module by a high
level driver such as SCPI or VXI
plug&play
. Changing values in registers
with register writes will confuse the driver because it has a record of the
register states after the last command executed by the driver. The driver
record of register states is not updated by a register write, only the register
value is changed. Therefore, the driver will not know the module state.
However, the SCPI driver will re-sync to the hardware if a
CLOSe?
query
is executed.
The following example program was developed with the ANSI C language
using the Agilent VISA extensions. The program was written and tested in
Microsoft
®
Visual C++ but should compile under any standard ANSI C
compiler.
To run the program you must have the Agilent SICL Library, the Agilent
VISA extensions, and an Agilent 82340 or 82341 GPIB module installed
and properly configured in your PC. An Agilent E1406 Command Module
is required.
Parameter
Default
Value
Description
ARM:COUNt
1
Number of scanning cycles is one.
TRIGger:SOURce
IMM
Advances through a scanning list
automatically.
INITiate:CONTinuous
OFF
Continuous scanning disabled.
OUTPut[:STATe]
OFF
Trigger output from EXT, TTL, or ECL
sources is disabled.
[ROUTe:]FUNC
WIRE2
Operating mode is set to WIRE2 at power-
on. This mode is NOT changed by *RST.
[ROUTe:]SCAN:MODE
NONE
Channel list is not set up.
[ROUTe:]SCAN:PORT
NONE
Analog bus connections are disabled from
channels.
Summary of Contents for E8462A
Page 2: ...x...
Page 10: ...8 Notes...
Page 11: ...9 Notes...
Page 12: ...10 Notes...
Page 15: ...Configuring the Agilent E8462A Multiplexer 13 Figure 1 1 Agilent E8462A Simplified Schematic...
Page 68: ...66 Using the Multiplexer...
Page 104: ...102 Agilent E8462A Relay Multiplexer Command Reference...
Page 118: ...116 Agilent E8462A Relay Multiplexer Command Reference...
Page 126: ...124 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4 Notes...
Page 146: ...144 Register Based Programming...
Page 156: ...154 Index...