Chapter 7
243
Digital Signal Interface Module
Operating the N5102A Module in Input Mode
Operating the N5102A Module in Input Mode
This section shows how to set the parameters for the N5102A module using the signal generator UI in the
input direction. Each procedure contains a figure that shows the softkey menu structure for the interface
module function being performed.
Refer to
“Connecting the Clock Source and the Device Under Test” on page 230
and configure the test
setup.
Accessing the N5102A Module User Interface
All parameters for the N5102A module are set with softkeys on the ESG signal generator.
Press
Aux Fctn
>
N5102A Interface
.
This accesses the UI (first-level softkey menu shown in
) that is used to configure the digital
signal interface module. Notice the graphic, in the ESG display, showing a setup where the N5102A module
is generating its own internal clock signal. This graphic changes to reflect the current clock source selection.
Figure 7-15
First-Level Softkey Menu
Line is grayed out until the N5102A module interface is turned on
Internal clock going to the DUT
Summary of Contents for E4428C
Page 22: ...Contents xxii ...
Page 224: ...200 Chapter 4 Basic Digital Operation Creating and Using Bit Files ...
Page 228: ...204 Chapter 5 AWGN Waveform Generator Configuring the AWGN Generator ...
Page 229: ...205 6 Analog Modulation ...
Page 276: ...252 Chapter 7 Digital Signal Interface Module Operating the N5102A Module in Input Mode ...
Page 286: ...262 Chapter 8 Bluetooth Signals Turning On a Bluetooth Signal ...
Page 330: ...306 Chapter 9 BERT Verifying BERT Operation ...
Page 366: ...342 Chapter 10 CDMA Digital Modulation IS 95A Modulation ...
Page 394: ...370 Chapter 12 Multitone Waveform Generator Applying Changes to an Active Multitone Signal ...
Page 468: ...444 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Frame Structures ...
Page 667: ...643 18 Troubleshooting ...
Page 700: ...Index 676 Index ...