Chapter 18
651
Troubleshooting
Basic Signal Generator Operations
Incorrect List Sweep Dwell Time
If the signal generator does not dwell for the correct period of time at each sweep list point, follow these
steps:
1. Press
Sweep/List
>
Configure List Sweep
.
This displays the sweep list values.
2. Check the sweep list dwell values for accuracy.
3. Edit the dwell values if they are incorrect.
NOTE
The effective dwell time at the RF OUTPUT connector is the sum of the value set for the
dwell plus processing time, switching time, and settling time. This additional time added to
the dwell is generally a few milliseconds. The TTL/CMOS output available at the
TRIG OUT connector, however, is asserted high only during the actual dwell time.
If the list dwell values are correct, continue to the next step.
4. Observe if the
Dwell Type List Step
softkey is set to Step.
When Step is selected, the signal generator will sweep the list points using the dwell time set for step
sweep rather than the sweep list dwell values.
To view the step sweep dwell time, follow these steps:
a. Press
Configure Step Sweep
.
b. Observe the value set for the
Step Dwell
softkey.
List Sweep Information is Missing from a Recalled Register
List sweep information is not stored as part of the instrument state in an instrument state register. Only the
current list sweep is available to the signal generator. List sweep data can be stored in the instrument catalog.
For instructions, see
Summary of Contents for E4428C
Page 22: ...Contents xxii ...
Page 224: ...200 Chapter 4 Basic Digital Operation Creating and Using Bit Files ...
Page 228: ...204 Chapter 5 AWGN Waveform Generator Configuring the AWGN Generator ...
Page 229: ...205 6 Analog Modulation ...
Page 276: ...252 Chapter 7 Digital Signal Interface Module Operating the N5102A Module in Input Mode ...
Page 286: ...262 Chapter 8 Bluetooth Signals Turning On a Bluetooth Signal ...
Page 330: ...306 Chapter 9 BERT Verifying BERT Operation ...
Page 366: ...342 Chapter 10 CDMA Digital Modulation IS 95A Modulation ...
Page 394: ...370 Chapter 12 Multitone Waveform Generator Applying Changes to an Active Multitone Signal ...
Page 468: ...444 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Frame Structures ...
Page 667: ...643 18 Troubleshooting ...
Page 700: ...Index 676 Index ...