Chapter 6
213
Analog Modulation
Configuring the LF Output
Configuring the LF Output with an Internal Modulation Source
In this example, the internal FM modulation is the LF output source.
NOTE
Internal modulation (
Internal Monitor
) is the default LF output source.
Configuring the Internal Modulation as the LF Output Source
1. Press
Preset
.
2. Press the
FM/
Φ
M
hardkey.
3. Press
FM Dev
>
75
>
kHz
.
4. Press
FM Rate
>
10
>
kHz
.
5. Press
FM Off On
.
You have set up the FM signal with a rate of 10 kHz and 75 kHz of deviation. The
FM
annunciator is
activated indicating that you have enabled frequency modulation.
Configuring the Low Frequency Output
1. Press the
LF Out
hardkey.
2. Press
LF Out Amplitude
>
3
>
Vp
.
3. Press
LF Out Off On
.
You have configured the LF output signal for a 3 volt sine wave (default wave form) output which is
frequency modulated using the
Internal Monitor
source selection (default source).
Summary of Contents for E4428C
Page 22: ...Contents xxii ...
Page 224: ...200 Chapter 4 Basic Digital Operation Creating and Using Bit Files ...
Page 228: ...204 Chapter 5 AWGN Waveform Generator Configuring the AWGN Generator ...
Page 229: ...205 6 Analog Modulation ...
Page 276: ...252 Chapter 7 Digital Signal Interface Module Operating the N5102A Module in Input Mode ...
Page 286: ...262 Chapter 8 Bluetooth Signals Turning On a Bluetooth Signal ...
Page 330: ...306 Chapter 9 BERT Verifying BERT Operation ...
Page 366: ...342 Chapter 10 CDMA Digital Modulation IS 95A Modulation ...
Page 394: ...370 Chapter 12 Multitone Waveform Generator Applying Changes to an Active Multitone Signal ...
Page 468: ...444 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Frame Structures ...
Page 667: ...643 18 Troubleshooting ...
Page 700: ...Index 676 Index ...