Power Meter Operation
2
E4418B Power Meter User’s Guide
63
Setting Offsets
Setting Channel Offsets
The power meter can be configured to compensate for a signal loss or gain
in your test setup (for example, to compensate for the loss of a 10 dB
attenuator). This gain or loss, referred to as a channel offset, is added to
the measured power before the result is displayed.
Offsets are entered in dB. The allowable range of values is –100 dB to
+100 dB. A positive value compensates for a loss and a negative value
compensates for a gain.
If either a channel or a display offset is set, then “Ofs” is displayed.
To enter a channel offset which compensates for a loss or gain:
1
Press
.
2
Press
.
3
Press
. The power meter displays the offset value in a pop- up
window. Modify this offset (see below) as desired.
•
Use
or
to modify the digit on which the cursor is
currently positioned.
•
Use
or
to move to other digits.
4
To confirm your choice, press
.
5
is automatically set to “On” when a value is entered
using
.
Channel offsets can be disabled and re- enabled simply by pressing
.
Input Settings
Offset
dB
Offset Off On
Offset
Offset Off On
Summary of Contents for E4418B
Page 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Page 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...