Power Meter Operation
2
E4418B Power Meter User’s Guide
31
1
Press
.
2
Connect the power sensor to the POWER REF output.
3
Press
to calibrate the power meter. During calibration the wait symbol
is displayed. (The power meter automatically turns on the POWER REF
output.)
Example
To calibrate the power meter with an Agilent E- Series power sensor.
•
Press
.
•
Connect the power sensor to the POWER REF output.
•
Press
.
Calibration Procedure using Agilent 8480 Series power
sensors and N8480 Series Power Sensors with Option CFT
The following procedure describes how you calibrate the power meter with the
Agilent 8480 Series power sensors and N8480 Series power sensors with
Option CFT.
Cal
Cal
N O T E
V8486A and W8486A sensors
For most 8480 Series sensors, the correct (A type or D type) linearity correction table is
automatically selected. The linearity-type field on the measurement screen is grayed out.
Under the gray-out condition, the selected linearity correction table that is displayed on
the screen is irrelevant to the correct linearity correction table that is automatically
selected by power meter firmware.
However, for the V8486A and W8486A sensors (
V8486A and W8486A sensors only
,) the
automatic selection must be overridden by selecting the D type linearity. Subsequent
connection of another A type sensor while the D type is selected will result in a warning
message stating that “Linearity Override May be Required”.
To select the linearity type to be applied:
Press
,
,
,
, or
.
Tables Linearity ATyp
DTyp
Summary of Contents for E4418B
Page 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Page 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...