![Agilent Technologies E4418B User Manual Download Page 192](http://html.mh-extra.com/html/agilent-technologies/e4418b/e4418b_user-manual_2868035192.webp)
5
Specifications
168
E4418B Power Meter User’s Guide
For Agilent N8480 Series power sensors
Manual filter, 10 dB decreasing power step:
Table 5-6
Settling Time
Auto filter, default resolution, 10 dB decreasing power step,
normal and x2 speed modes:
Number of Averages
1
2
4
8
16
32
64
128
256
512
1024
Settling Time (s)
(Normal Mode)
0.15
0.2
0.3
0.5
1.1
1.9
3.4
6.6
13
27
57
Response Time (s)
(x2 mode)
0.15
0.18
0.22
0.35
0.55
1.1
1.9
3.5
6.9
14.5
33
10 dB
10 dB
10 dB
20 dB
Minimum dBm
Power
150 ms
6.6 s
1 s
200 ms
Settling
Times
Sensor
Dynamic
Range
Maximum dBm
Normal Mode
Typical
13.5 s
5 dB
10 dB
10 dB
10 dB
20 dB
Power
150 ms
3.6 s
400 ms
180 ms
Settling
Times
Sensor
Dynamic
Range
Maximum dBm
Minimum dBm
x2 Mode
Typical
6.6 s
5 dB
Summary of Contents for E4418B
Page 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Page 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...