Power Meter Operation
82
E4418B Power Meter User’s Guide
To set the serial interface parameters:
1
Press
,
,
,
. The
current settings for
,
,
, and
are
shown under the softkeys.
2
To change the baud rate, press
. The power meter displays
the baud rate in a pop- up window. Change the setting by using the
and
keys to run through the selection of baud rates
available. To confirm your choice, press
.
3
To change the word size, press
. The power meter displays
the word size in a pop- up window. Change the setting by using any of
the arrow keys to toggle between a word size of seven or eight. To
confirm your choice, press
.
4
To change the number of stop bits, press
. The power meter
displays the number of stop bits in a pop- up window. Change the
setting by using any of the arrow keys to toggle between a setting of
one or two. To confirm your choice, press
.
5
To change the parity type, press
. The power meter displays the
parity type in a pop- up window. Change the setting by using any of the
arrow keys to select from EVEN, ODD, ZERO, ONE, or NONE. To
confirm your choice, press
.
6
Press
to access page 2 of 2. This page allows you to set echo off
and on and gives access to the pacing menu.
7
Press
to toggle between off and on. When echo is on, all received
characters will be transmitted (echoed) back to the sender.
8
Press
to access the pacing menu which allows you to enable or
disable various software and hardware pacing options. Pacing will
prevent receiver overrun errors and is generally only required when
large amounts of data are being transferred (for example, cal tables).
9
Press
to toggle between enabling and disabling the Xon/Xoff
transmitter software handshake. When enabled “Xon” is highlighted,
otherwise “None” is highlighted.
10
Press
to toggle between enabling and disabling the Xon/Xoff
receiver software handshake. When enabled “Xon” is highlighted,
otherwise “None” is highlighted.
11
Press
and use any of the arrow keys to select one of the
following from the pop- up menu:
OFF - Sets the RTS signal line permanently low.
Remote Interface
Configure Interface
Serial
Baud rate
Word size
Stop bits
Parity
Baud rate
Enter
Word size
Enter
Stop bits
Enter
Parity
Enter
Echo
Pacing
Tx Pacing
Rx Pacing
RTS/CTS
Summary of Contents for E4418B
Page 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Page 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...